MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problems Device Sensing Problem (2917); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2020
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.The unique identifier was not known at the time of reporting.Concomitant medical products: product id: 9734887xom, serial/lot #: (b)(4).Product id: 9733533, serial/lot #: (b)(4), ubd: 02-feb-2022.Product id: 9660651, serial/lot #: (b)(4), ubd: 02-feb-2022.No parts have been received by the manufacturer for evaluation.) the manufacture date was not known at the time of reporting.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system.It was reported outside of a procedure prior to a case that the emitter box would not recognize trackers and that the em quadcut 4.3 mm may have been faulty so the igs was not used.No patient was present at the time of the event.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.It was suspected that the issue was caused by either human error or an issue with the electromagnetic navigation interface unit.The system then passed the system checkout and was found to be fully functional.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was reported that two ports on the emitter box were not operational.To complete the subsequent procedure, the site used the other two ports on the unit.It was noted that replacement of the emitter box restored functionality.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11010981
MDR Text Key221829258
Report Number1723170-2020-03278
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2020
Initial Date FDA Received12/15/2020
Supplement Dates Manufacturer Received12/18/2020
03/16/2021
Supplement Dates FDA Received01/12/2021
03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10."
-
-