• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD 2000 SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD 2000 SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A2000
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the swivel lock of the mayfield 2000 skull clamp does not lock the rocker arm tightly.It has some play or movement in the rocker arm when in the locked position.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
 
Manufacturer Narrative
Unique device identifier (udi): (b)(4).The reported complaint was confirmed from the evaluation of the returned mayfield skull clamp.Evaluation showed that the hex bushing is worn and the lock still moves after unit is lock down.The unit needs pm, repair, and replacement of worn/damaged parts.This device exceeds its expected life of 7 years ( lot code: 994: manufactured in 1999).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD 2000 SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key11015272
MDR Text Key226102060
Report Number3004608878-2020-00732
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K932807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2000
Device Lot Number994
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Date Manufacturer Received12/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-