MEDTRONIC HEART VALVES DIVISION AVALUS AORTIC TISSUE VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE
|
Back to Search Results |
|
Model Number 40023 |
Device Problem
Inadequacy of Device Shape and/or Size (1583)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that prior to implant, the surgeon sized the patient with a 23mm sizer and noted it was a good fit.A 23mm bioprosthetic valve was used, however would not seat down to the annulus.The surgeon elected to remove the valve and successfully implanted a 23mm non medtronic valve in its place.No additional adverse patient effects were reported. .
|
|