Product analysis: upon receipt at medtronic's quality laboratory, the valve was discolored showing evidence of blood contact.All leaflets were flexible, intact, and in the closed position.No leaflet damage was observed.All commissures were intact and no commissure damage was noted.The valve sewing ring was measure and found acceptable within the 25mm size valve specification.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.This valve was re-assessed for size confirmation and confirmed to meet requirements.Based on the information available, the root cause of this complaint cannot be determined.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that during the implant of this 25mm bioprosthetic valve, the sutures were set into the sewing ring and valve pulled down, but the valve could not pass through the sinotubular junction (stj).The physician re-sized and confirmed both ends of the sizer read 25mm.The 25mm was retrieved.A 23mm bioprosthetic valve of the same model was implanted.No additional adverse patient effects were reported.
|