Model Number 60-05-65 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 erc tubing clamp / 620mm.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.The clamp was functional but noisily.The device will be returned to the manufacturer site for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 erc tubing clamp / 620mm was stuck during priming and displayed error codes.The user did not remember which error codes had appeared.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: the device was returned to the manufacturer site and the reported failure was confirmed.Several parts were replaced to solve the issue and the unit was then shipped back to the customer.Based on the investigation performed for similar cases, the most likely root cause of the reported event is assigned to fluid ingress and use conditions (i.E.Extensive exposure to liquid, e.G.During cleaning).
|
|
Event Description
|
See initial report.
|
|