It was reported via clinic notes that the patient's settings were disabled for an unknown reason.The programming history data for the device was analyzed and though no instances of the burst watchdog error were seen, based on report of the burst watchdog error being seen on the tablet, and evidence of the patient's autostimulation mode output current and magnet mode output currents were set equal to each other, it is likely that a burst watchdog disablement occurred.After re-programming the device so that the autostim output current was less than the magnet output current, it was confirmed within the programming history that the vns therapy was re-enabled.However, burst watchdog disablement cannot ultimately be confirmed since there was no error code reported to be seen upon interrogation of the patient initially when disablement was identified.The burst watch dog timeout is a known issue occurs when a rare combination of circumstances are present, one of which is when the autostim output current is programmed greater or equal to the magnet output current.No additional relevant information has been received to date.
|
The programming history data for the device was analyzed and confirmed that a burst watchdog disablement had occurred.Due to unintended behavior in the model 106 aspire sr generator software, the generator can stop delivering stimulation unexpectedly.This event can only occur when a rare combination of circumstances are present, one of which is when the autostim output current is programmed greater or equal to the magnet output current.In this case the autostim output current mode was programmed equal to the magnet output current.Once the autostim output current was decreased to an output current less than the magnet mode output current, the vns pulse therapy was enabled.No additional relevant information has been received to date.
|