TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problems Battery Problem (2885); Output Problem (3005)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2020
Event Type  malfunction  
Manufacturer Narrative
Per the field service representative (fsr) the user facility's biomedical engineer stated that there were power issues in the operating room.
 
Event Description
It was reported that the heart lung machine (hlm) had a battery error issue.No other details regarding the nature of the event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.The product surveillance technician duplicated the reported complaint.He observed the voltage to be too low that it was unable to power on the fans of the power supply's.The voltage was reading 0.112 volts direct current (vdc), specification is 24.5 vdc +/- 5%.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Additional information received that the surgical procedure was completed successfully.No adverse consequences to the patient.
 
Manufacturer Narrative
The field service representative (fsr) was unable to duplicate any issues.After consulting further with the user facility's biomedical technician it was determined that a bad power outlet caused the heart lung machine (hlm) to go onto battery power without the user knowing.There are no issues with the battery or the hlm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11023400
MDR Text Key224410994
Report Number1828100-2020-00480
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000656
UDI-Public(01)00886799000656(11)190731
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-