• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported while testing for sars cov-2 a false positive result was obtained.The patients were retested 3 times on the veritor and the results were negative.No additional information was provided by the customer.Eua#: (b)(4).
 
Manufacturer Narrative
H.6.Investigation: bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples if applicable.The investigation did not find a root cause for the false positive results that were observed.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.H3 other text : see h.10.
 
Event Description
It was reported while testing for sars cov-2 a false positive result was obtained.The patients were retested 3 times on the veritor and the results were negative.No additional information was provided by the customer.Eua#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
MDR Report Key11023415
MDR Text Key223181559
Report Number1119779-2020-01237
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number256082
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-