H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for.During evaluation, the reported issue of an image drop out is was unconfirmed.The ultrasound image was clear and showed no issues of working improperly.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|