• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: bi71000162, serial/lot #: unknown.A medtronic representative went to the site to test the system.Testing revealed that tray 8 was bad, or has a bad cell.Replaced tray 8.Voltage remained stable after.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information from a manufacturer representative (rep) regarding a device being used outside of a procedure.It was reported that the o2 battery is draining faster than normal.There was no patient harm or additional complications reported or anticipated as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key11023724
MDR Text Key222074276
Report Number3004785967-2020-01523
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/07/2020
Initial Date FDA Received12/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse