U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX CUSTOM PACK; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX CUSTOM PACK; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX-XRX00660
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
Udi: not required for product code.Implanted date: device was not implanted.Explanted date: device was not explanted.Address: (b)(6) hospital.Occupation: clinical engineer.Pma/510(k)- k130520.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.A review of the device history record and product-release judgement record of the involved product code/lot# combination was conducted with no findings.Review of the shipping inspection record of the product in which the involved fiber lot was used confirmed that there was no anomaly in the gas exchange performance test and pressure drop test results.The test results satisfied the factory's control criteria.Ifu states: do not use this product for a period in the excess of six hours.Excessive use for over six hours may lead to plasma leak and thrombi formation, which may compromise the gas exchange performance.A phenomenon called wet lung may occur when water condensation occurs inside fibers of microporous membrane oxygenators with blood flowing exterior to the fibers.This may occur when oxygenators are used for a longer period of time.If water condensation and/or a decrease in pao2 and/or an increase in paco2 is noted during extended oxygenator use, briefly increasing the gas flow rate may improve the performance.Increase gas flow rate, to 20 l/min for 10 seconds.Do not repeat this flushing technique, even if oxygenator performance is not improved.It is likely that due to a change in the blood properties for some reason, a surface-active substance may be generated in blood.This may lead the balance of the surface tension between the gas and blood which is kept at the micro pores on the surface of the fibers to be upset, and the fibers got hydrophilized, resulting in plasma leak; the pressure inside the oxygenator module raised by some factors, such as clotting, may cause the pressure applied from the blood channel to the gas channel to get increased.This may create the circumstances where force to push the blood corpuscle components out into the gas phase may increase and plasma component could easily leak out through the micro pores on the surface of the fibers to the gas channel.However with no return of the actual sample, the exact cause of the reported event cannot be definitively determined based on the available information.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
 
Event Description
The user facility reported that the involved capiox custom pack was used during the re-ope bentall procedure.Fio2 around 80% and pao2 around 190 in seven hours of circulation.In eight hours, another oxygenator was added to cope with the drop in oxygenation performance.The procedure outcome was not reported.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX CUSTOM PACK
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
theresa mussaw
reg. no. 2243441
265 davidson ave suite 320
somerset, NJ 08873
8002837866
MDR Report Key11026267
MDR Text Key227998322
Report Number9681834-2020-00253
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberCX-XRX00660
Device Lot Number200715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-