• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770552
Device Problems Loss of Power (1475); Unexpected Shutdown (4019)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
The device, even when fully charged, suddenly turns off and happened during a procedure.
 
Event Description
The device, even when fully charged, suddenly turns off and happened during a procedure.
 
Manufacturer Narrative
Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of battery shuts down was confirmed.The rite 8 site was visually inspected upon receipt and showed no physical damage.The reported problem was confirmed, the equipment shuts down after removal of the power supply and only stays on if it is connected to power.The main cause of the problem was a premature failure of the lithium battery that does not charge.The battery has the number of cycles below the critical value (number of cycles: 42) but does not provide the correct voltage for proper operation of the equipment.A history review of serial number dycxad518 showed no other similar product complaint(s) from this serial number.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key11028768
MDR Text Key222643994
Report Number3006260740-2020-20992
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098093
UDI-Public(01)00801741098093
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Remedial Action Notification
Type of Report Initial,Followup
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2021
Initial Date Manufacturer Received 11/25/2020
Initial Date FDA Received12/17/2020
Supplement Dates Manufacturer Received06/24/2021
Supplement Dates FDA Received06/30/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2243072-10/11/2019-013-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-