• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0480-01
Device Problem Restricted Flow rate (1248)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2020
Event Type  malfunction  
Manufacturer Narrative
Complete event site address: (b)(6).The device has not been returned to the manufacturer, so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id #: (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console would not run and generated a check iab catheter alarm.The customer determined there was no visible abnormality on the iab, but the alarm continued.The iab was replaced, and the alarm stopped.There was no patient harm, or adverse event reported.
 
Manufacturer Narrative
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Reference complaint # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console would not run and generated a check iab catheter alarm.The customer determined there was no visible abnormality on the iab, but the alarm continued.The iab was replaced and the alarm stopped.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The product was returned with the membrane completely unfolded and traces of blood on the exterior of the catheter.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The iab was placed on the cs300 pump and pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported event cannot be confirmed by the evaluation.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint #(b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console would not run and generated a check iab catheter alarm.The customer determined there was no visible abnormality on the iab, but the alarm continued.The iab was replaced and the alarm stopped.There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11029540
MDR Text Key222338598
Report Number2248146-2020-00673
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567106564
UDI-Public10607567106564
Combination Product (y/n)N
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/08/2022
Device Model Number0684-00-0480-01
Device Catalogue Number0684-00-0475
Device Lot Number3000095711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Date Manufacturer Received04/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age70 YR
Patient Weight63
-
-