Model Number BI70002000 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Aware date of initial mdr reported incorrectly.Value should equal 02 dec 2020." if information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: bi71000860, serial/lot #: (b)(4).A representative went to the site to preform system check out.It was noted that the generator was not starting.They replaced the power conversion and that resolved the issue.The power conversion board was returned and it was noted that they were able to confirm the reported issue.There were two transistors that were found to be shorted.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
(b)(6) 2020 (b)(4) (rep) medtronic received information regarding an imaging system that was used outside of procedure.It was reported that the power conversion board is damaged and needs to be replaced.No patient present.
|
|