• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SANMINA -SCI SYSTEMS O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SANMINA -SCI SYSTEMS O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Device Problems Calibration Problem (2890); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2020
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative visited the site to perform a system checkout.The product passed all tests and was performing as intended.No hardware parts were replaced.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information from a manufacturer representative regarding an imaging system outside of a procedure.It was reported that the site was getting ready for a case later in the week.They went to calibrate the system and when they went onto the screen and it had remained stuck in preparing.No patient was present.Additional information was received: it was reported that it was stuck in the gain calibration screen.They said the system gave a warning about gain calibrations and when they went to that tab and hit okay it was unresponsive in that screen overnight.Technical services had them reboot the system and it was able to boot up normally and take a test 2d image without issue.It was recommended to perform regular gain calibrations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key11031853
MDR Text Key223621637
Report Number3006544299-2020-00498
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public00763000132378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Date Manufacturer Received12/11/2020
Date Device Manufactured09/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-