• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER320
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to obtain additional information and to retrieve the device.To date, no additional information has been received and no device has been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent: please clarify " the jaw overlaps", was the jaw misaligned ? if other, please specify.Please provide the status of the device(s) as it has not been received for analysis.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure, the clips "get out like x shape from the jaw overlaps.Surgeon used oversewing technique instead of clip applier.There was no patient consequence.
 
Manufacturer Narrative
(b)(4) date sent: 1/4/2021 h2: additional information received: please clarify " the jaw overlaps," was the jaw misaligned ? if other, please specify.No.I mean the clips take x shape when it get out from the jaw.H6: medical device problem code: is only failure to form staple (a050704).The device did not have a damaged jaw so mechanical problem (a05) medical device problem code should be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key11033722
MDR Text Key222699331
Report Number3005075853-2020-06974
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER320
Device Catalogue NumberER320
Device Lot NumberU4063Y
Was Device Available for Evaluation? No
Date Manufacturer Received12/28/2020
Patient Sequence Number1
-
-