Model Number CD3371-40Q |
Device Problem
Failure to Disconnect (2541)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
During device replacement due to normal eri, the left ventricular (lv) lead was unable to be removed from the device header.The set screw was unable to be loosened.Upon lv lead removal attempt the lead broke, and pin remained in the header.The event was resolved by explanting and replacing the lv lead, and device due to normal eri.The patient was in stable condition.Related to manufacturer report number: 2017865-2020-23111.
|
|
Manufacturer Narrative
|
The reported event of failure to disconnect was confirmed in the lab.Interrogation of the device revealed the device was at the elective replacement indicator when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Pacing, sensing, impedance, hv output, patient notifier was tested and no anomaly was detected.A partial lead was stuck in the lv bore of the device.This partial lead was unable to be removed due to the lvtip set screw being stripped.Remains of septum material was found in the lvtip set screw hex cavity.The septum material in the hex cavity prevented full insertion of the torque driver and was the cause of the field event.
|
|
Search Alerts/Recalls
|