Model Number 71440008
Device Problems Material Separation (1562); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2020
Event Type  malfunction  
Event Description
It was initially reported that this instrument was not fit for use.No case involved.However, based on a preliminary assessment conducted by the manufacturer, it was determined that this event needs to be re-classified as reportable as the knob was loose.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device has scratches and nicks on the surface.The device was manufactured in 2011 and shows signs of extensive use.A functional evaluation of the returned device could confirm the stated failure mode.The adjustment knob is loose, rendering the device inoperable.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII DCF VARIABLE SZ GUIDE RT
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11037903
MDR Text Key222349600
Report Number1020279-2020-07577
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010610393
UDI-Public03596010610393
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440008
Device Catalogue Number71440008
Device Lot Number11DM02276
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Date Manufacturer Received01/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-