• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem No Audible Alarm (1019)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418); Sweating (2444)
Event Date 12/06/2020
Event Type  Injury  
Manufacturer Narrative
Abbot diabetes care product quality engineering (pqe) investigated the alarm-3 (missing high or low glucose ) complaint.The serial number of the freestyle libre 2 sensor associated with this complaint is unknown.During investigation of the track and trend review related to alarm-3 cases in (b)(6) 2020 this failure mode was identified and is a known manufacturing issue that impacts libre 2 sensors manufactured at flex buffalo grove (fbg).As the serial number of the sensor associated with this complaint is unknown, pqe is unable to determine if the sensor is impacted by this manufacturing issue.Outside of the known manufacturing issue, the available tripped trend reports for libre 2 sensor and alarm-3 have been reviewed.The review did not identify any additional trends that would indicate a product related issue related to this complaint.If the product is returned, the case will be re-opened, and a physical investigation will be performed.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported missing the low glucose alarm while wearing the adc freestyle libre 2 sensor.On (b)(6) 2020, as a result, the customer experienced cold sweats and lost consciousness.The customer was treated with an injection of glucagon by the customer's partner.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
erica frank
1360 south loop road
alameda, CA 94502-7001
5104242454
MDR Report Key11041097
MDR Text Key222619327
Report Number2954323-2020-14031
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/16/2020
Initial Date FDA Received12/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
Patient Weight90
-
-