• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.201
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Reporter¿s phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported from (b)(6) that during post-surgery, it was discovered that the button on the battery handpiece/modular device was jammed.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: this device was returned for evaluation.A visual and functional assessment was performed which determined that the device passed all pre-repair diagnostic assessments and no failures were identified.Therefore, the reported condition was not confirmed, and an assignable root cause was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11041429
MDR Text Key245043557
Report Number8030965-2020-09826
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819977815
UDI-Public07611819977815
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Date Manufacturer Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-