Model Number 1456Q/86 |
Device Problems
Difficult to Remove (1528); Separation Failure (2547)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented for a routine implant.It was observed during the procedure that after the physician had tied a knot at the distal end of the left ventricular lead in the coronary sinus, the guidewire could not be removed from the lead.The lead was pulled out and exchanged.The patient was stable.
|
|
Manufacturer Narrative
|
The damage found was sustained during procedure.The lead was otherwise normal.The reported event of unable to remove the guidewire was confirmed.A complete lead was returned for analysis.Visual analysis noted a knot at the distal tip region of the guidewire.The cause of the reported event was isolated to the damaged found on the associated guidewire.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies with the exception of procedural damage.
|
|
Search Alerts/Recalls
|