Model Number G148
Device Problems Off-Label Use (1494); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Sepsis (2067); No Code Available (3191)
Event Date 10/12/2020
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product analysis indicates that the allegation against the device was not confirmed.The device was analyzed, passed all the baseline tests and exhibited normal device functions.If information is provided in the future, a supplemental report will be issued.(b)(4).
 
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) had an infection with sepsis.A revision was performed and the crt-d was explanted.Additional information received indicates that the crt-d was connected to the temporary lead for temporary pacing.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11042409
MDR Text Key222688021
Report Number2124215-2020-25686
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/01/2020
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number193708
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received12/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age76 YR
-
-