• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. FA COBLATOR II CONTROLLER (120V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. FA COBLATOR II CONTROLLER (120V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 13546-01
Device Problems Overheating of Device (1437); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that, the "coblator ii controller" was noisy and overheated.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10, h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A visual inspection observed the unit is missing two feet, the volume knob grinds when spun, and the lid is damaged.A functional evaluation revealed volume does not function.There were no signs of overheating issues.The unit was opened and found no issues.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FA COBLATOR II CONTROLLER (120V)
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11043023
MDR Text Key239213482
Report Number3006524618-2020-01111
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
PMA/PMN Number
K030108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number13546-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Date Manufacturer Received01/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-