MPRI ATTAIN PERFORMA S MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 459888
Device Problems High impedance (1291); Capturing Problem (2891); Impedance Problem (2950); High Capture Threshold (3266)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 11/30/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced severe hypotension during hemodialysis treatment which required hospitalization and administration of fluids.Pacing impedance and thresholds were noted to rise around that time with impedance subsequently reaching a high undefined range and thresholds reaching a high range.The lead was reprogrammed with electrical measurements in normal range.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA S MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11045991
MDR Text Key222646130
Report Number2649622-2020-24999
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601871
UDI-Public00643169601871
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/03/2020
Device Model Number459888
Device Catalogue Number459888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
W4TR03 CRT-P, 5076-45 LEAD, 5076-52 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age72 YR
-
-