• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP Back to Search Results
Catalog Number A2114
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the ratchet plunger knob mechanism of the mayfield infinity xr2 skull clamp (a2114 ) was broken.It is unknown if there was patient involvement, or if the device failure led to patient injury, or surgical delay.
 
Manufacturer Narrative
Unique device identifier (udi): (b)(4).Device history record (dhr) - the dhr documentation showed no abnormalities related to the reported failure.The mayfield skull clamp was returned for evaluation: failure analysis: the unit was received with the left and right pawls worn and the lock was difficult to lock/unlock.Unit required general maintenance and cleaning.Root cause: the reported complaint was confirmed via inspection.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Type of Device
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key11046369
MDR Text Key223789741
Report Number3004608878-2020-00748
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K130389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2020
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received12/23/2020
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-