Catalog Number A2114 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the ratchet plunger knob mechanism of the mayfield infinity xr2 skull clamp (a2114 ) was broken.It is unknown if there was patient involvement, or if the device failure led to patient injury, or surgical delay.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).Device history record (dhr) - the dhr documentation showed no abnormalities related to the reported failure.The mayfield skull clamp was returned for evaluation: failure analysis: the unit was received with the left and right pawls worn and the lock was difficult to lock/unlock.Unit required general maintenance and cleaning.Root cause: the reported complaint was confirmed via inspection.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|
Event Description
|
N/a.
|
|