Model Number 1304.15.140
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Unspecified Infection (1930)
Event Date 12/18/2020
Event Type  Injury  
Manufacturer Narrative
We are currently investigate the case at the best of our possibilities.We will send a final incident report once the investigation will be concluded.
 
Event Description
Revision surgery of a smr reverse prosthesis (with customized glenoid component found to be loose) due to deep infection performed on (b)(6) 2020.During the revision surgery, all the components previously implanted during the previous surgery performed on (b)(6) 2020 have been explanted.Following, the product details of the explants: smr cementless finned stem, code 1304.15.140, lot# 1917902, ster.2000046; smr reverse humeral body, code 1352.20.010, lot# 2011653, ster.2000274; smr reverse liner standard, code 1360.50.010, lot# 19at1km, ster.1900360; smr connector small std, code 1374.15.310, lot# 1916953, ster.1900376; smr eccent.Glenosphere ø 36mm, code 1376.09.031, lot# 1912500, ster.2000020; bone screw d.4,5 l.26mm, code 8431.15.026, lot# 1921132, ster.1900441; bone screw d.4,5 l.28mm, code 8431.15.028, lot# 1920934, ster.1900462; bone screw d.4,5 l.34mm, code 8431.15.034, lot# 1912668, ster.1900305; customized glenoid, code 9961714y57000, lot# 2017117, ster.2000292.According to the info reported by the complaint source, pathogen responsible for the infection is unknown and patient suffers for parkinson's disease.No implants implanted during the revision surgery.Event happened in (b)(6).
 
Event Description
Shoulder revision surgery of a smr reverse prosthesis performed on (b)(6), 2020, due to deep infection.It was reported that the site developed redness and swelling which had progressed to deep infection.According to the complaint source, the pathogen responsible for the infection is unknown.It was reported that the customized glenoid component was found to be loose, whereas the humerus was well fixed.All components previously implanted had been explanted.Following, the product details of the explants: · smr cementless finned stem, code 1304.15.140, lot# 1917902, ster.2000046; · smr reverse humeral body, code 1352.20.010, lot# 2011653, ster.2000274 - product not sold in the us; · smr reverse liner standard, code 1360.50.010, lot# 19at1km, ster.1900360 - product not sold in the us; · smr connector small std, code 1374.15.310, lot# 1916953, ster.1900376; · smr eccent.Glenosphere ø 36mm, code 1376.09.031, lot# 1921500, ster.2000020; · bone screw d.4,5 l.26mm, code 8431.15.026, lot# 1921132, ster.1900441; · bone screw d.4,5 l.28mm, code 8431.15.028, lot# 1920934, ster.1900462; · bone screw d.4,5 l.34mm, code 8431.15.034, lot# 1912668, ster.1900305; · customized glenoid, code 9617.14.Y57, lot# 2017117, ster.2000292.No implants were implanted during the revision surgery.Previous surgery was performed on (b)(6), 2020.Patient is a female, 77 years old.It was reported that she suffers of parkinson's disease.Event happened in new zealand.
 
Manufacturer Narrative
By checking the sterilization charts of the involved lot #s, no pre-existing anomaly was found on the overall number.Therefore, all the products with these lot #s have been properly sterilized before being placed on the market.Explants analysis: the items involved were not available to be returned to limacorporate for further analysis.X-rays analysis: limacorporate received one x-ray referring to post-operative previous surgery.The x-ray received - exact date not known - has been evaluated by a medical consultant.Following, the medical consultant comments: "the image i think is post operative the revision procedure on the (b)(6).There are long standing changes in the antero-inferior glenoid, and obviously there has been at least one procedure prior to that revision.We do not have any other details but with 2 prior procedures the risk of infection is significantly increased.I do not see any features of the positioning of the implants that is concerning and at that stage there is no obvious bony change that would be consistent with infection.With the limited information available we conclude that the infection occurred in association with the revision procedure on the (b)(6).Explanting all prosthetic material is appropriate.There is no mention of a prostalac being used which would be a reasonable option".Considering that: · check of the sterilization charts highlighted no anomalies on the components manufactured with the same lot #s; · according to the medical consultant "with the limited information available we conclude that the infection occurred in association with the revision procedure on the (b)(6)."; we can state that the event was not product related.Pms data according to limacorporate pms data, revision rate of smr reverse prosthesis due to infection is (b)(4).Based on the root cause analysis performed and according to the relevant pms data, no corrective actions required for this specific case.Limacorporate will continue monitoring the market to promptly detect any further similar issue.Note: this is a final mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR CEMENTLESS FINNED STEM
Type of Device
FINNED HUMERAL STEMS - 80MM (TI6AL4V) DIA.14MM
Manufacturer (Section D)
LIMACORPORATE S.P.A.
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
MDR Report Key11046517
MDR Text Key223183908
Report Number3008021110-2020-00120
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
PMA/PMN Number
K101263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number1304.15.140
Device Lot Number1917902
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/18/2020
Initial Date FDA Received12/21/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received03/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-