According to the reporter, during the procedure, when trying a trial firing after opening the package prior to use, the counter decreased, but the clip was not loaded.A new product was opened and used to complete the case.When the sales rep received the product, the counter was found to be displayed as 07.There was no patient injury.
|
Additional information:g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument was received partially applied with one clip remaining, the clip counter was no longer active, and no visual abnormalities were observed with the instrument.Functionally, the instrument was applied to appropriate test media.The instrument cycled without binding.The clip advanced into the jaws, formed properly, and was held securely in place after full formation was ach ieved and the firing handle was released.The instrument did not load the next clip.Microscopic evaluation of the clip channel noted no clips were present, but the interlock would not engage.It was reported that the clips did not load properly into the jaws as expected.The reported issue was confirmed.The most likely cause could not be established from the information available.It was also reported that the digital reader did not work properly as intended.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|