Model Number 16422 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The srt found that during preventative maintenance (pm) of the device, the universal power supply board (upsb) was defective.The trimmer potentiometer adjustment screw on the ups battery module board would not adjust the charging voltage.It was spinning without making any charging voltage adjustments.The upsb board was replaced.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
The service repair technician (srt) reported that during testing of the device at the service center, the universal power supply (ups) battery module board was defective.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|