Model Number 3851
Device Problems Material Rupture (1546); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2020
Event Type  malfunction  
Event Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and moderately calcified mid left anterior descending artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During procedure, the mild tortuousity made it difficult to cross the lesion.Subsequently, the balloon ruptured upon second inflation at 4atm for 30 seconds when ballooning halfway into the lesion.The device was simply removed from the patient's body.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11052921
MDR Text Key223052360
Report Number2134265-2020-18246
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/26/2022
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0025923122
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/11/2020
Initial Date FDA Received12/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-