• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-150
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus (b)(4), it was found that there was the yellow stain and/or dirt inside of the light guide lens of the subject device.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device in this report was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Based upon the information from olympus india there was the possibility that the reported phenomenon was attributed to the moisture or foreign object invasion into the device from the air leakage part because there was air leakage at the control section.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11053012
MDR Text Key225929953
Report Number8010047-2020-10796
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170307553
UDI-Public04953170307553
Combination Product (y/n)N
PMA/PMN Number
K954451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberTJF-150
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-