• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problem Power Problem (3010)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
Per the fsr, the power supply was replaced and passed testing.The unit operated to the manufacturer's specifications.The suspect part will be sent back to the manufacturer for further evaluation.
 
Event Description
The field service representative (fsr) reported that during preventative maintenance (pm) of the device, the power supply failed.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Updated block: d10.During laboratory analysis, the product surveillance technician (pst) was unable to duplicate the reported complaint.He observed the power supply to meet all required specifications when tested three different times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11055495
MDR Text Key223411939
Report Number1828100-2020-00490
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000656
UDI-Public(01)00886799000656(11)171220
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2020
Initial Date FDA Received12/22/2020
Supplement Dates Manufacturer Received12/21/2020
01/18/2021
Supplement Dates FDA Received01/11/2021
02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-