Model Number 801763 |
Device Problem
Power Problem (3010)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the fsr, the power supply was replaced and passed testing.The unit operated to the manufacturer's specifications.The suspect part will be sent back to the manufacturer for further evaluation.
|
|
Event Description
|
The field service representative (fsr) reported that during preventative maintenance (pm) of the device, the power supply failed.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Updated block: d10.During laboratory analysis, the product surveillance technician (pst) was unable to duplicate the reported complaint.He observed the power supply to meet all required specifications when tested three different times.
|
|