Model Number ARD568350900
Device Problems Corroded (1131); Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/17/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided upon results of investigation.Device not returned to manufacturer.
 
Event Description
On (b)(6) 2020 getinge became aware of an issue with one of our surgical lights ¿ powerled.As it was stated paint peeling and rust was noticed on the device.No information about any injury has been provided, however we decided to report this case based on potential and in abundance of caution as any paint or rust particles falling from the device might led to contamination.
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
Getinge became aware of an issue with one of our surgical lights ¿ powerled.As it was stated paint peeling and rust was noticed on the device.No information about any injury has been provided, however we decided to report this case based on potential and in abundance of caution as any paint or rust particles falling from the device might led to contamination.It was established that when the event occurred, the surgical light did not meet its specification, since appearance of chipped paint and rust could be considered as technical deficiency, and in this way device contributed to event.We have not received information whether the device was being used for patient treatment at the time when the event occurred.When reviewing similar reportable events for the same device type, we have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.All maquet sas products comply with: iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.Iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.Paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.Disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).It was also established that the lack or degradation of the paint could be developing corrosion over time.The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid the collisions between devices.Visual inspections during the cleaning allow detecting the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint; nevertheless, the parts impacted by serious damage must be replaced.To avoid paint degradation and corrosion it is also recommended to respect the cleaning instructions avoiding spraying, high concentrations, prolonged exposure to detergents / disinfectants solutions, and to wipe with a dry cloth and to make sure no liquid residue is left on the device after cleaning.We believe that all remaining devices are performing correctly in the market.Given the circumstances and after our review of complaint ratio behavior of this nature, we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLED
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key11055862
MDR Text Key223161928
Report Number9710055-2020-00518
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberARD568350900
Device Catalogue NumberARD568350900
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-