MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0800-0543
Device Problems No Display/Image (1183); Display or Visual Feedback Problem (1184); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
The device return is anticipated, however; at the time of the report the devices have not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event Description
The customer reported that during a patient procedure, using a glidescope spectrum smart cable, the cable had to be held in a certain spot in order to get an image.The biomed isolated the issue to the smart cable.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Manufacturer Narrative
A replacement glidescope spectrum smart cable was provided to the customer and the smart cable used in the procedure was returned to verathon for evaluation.A verathon technical service representative evaluated the returned smart cable and was unable to confirm the "loss of image" issue.Since the customer's glidescope spectrum smart cable had already been replaced the returned smart cable was scrapped.No further investigation is required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE SPECTRUM SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
MDR Report Key11060800
MDR Text Key223809832
Report Number9615393-2020-00271
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0800-0543
Device Catalogue Number0800-0544
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Date Manufacturer Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-