There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Battery (voltage breaks down under load) defective, replaced.The housing is cracked in several places (probably due to a shock) defective, replaced.Motor connection cable (insulation broken) defective, replaced.Micromotor smr102752 tested, without errors.Supplied without contra-angle handpiece, power supply unit and foot control.Function test and test measurements without errors.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
|