This is filed to report peeled material.It was reported that during preparation of the steerable guide catheter (sgc), the dilator was inserted into the sgc, but it was noticed that a piece of the dilator had become frayed.The physician decided to not use the device and replace it.There was no clinically significant delay in the procedure.No additional information was provided.
|
The returned device analysis was unable to confirm the reported peeled/delaminated dilator tip; however, it was observed there was irregular appearance of the slanted/cut dilator tip.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other incidents reported from this lot.A cause for the reported peeled/delaminated issue could not be determined.However, the investigation identified a potential product quality issue due to irregular appearance of the slanted/cut dilator.The issue is being addressed per internal operation procedures.Abbott vascular will continue to trend the performance of these devices.
|