Product event summary: the data files were returned and analyzed.The data files confirmed receipt of the system notice 50030, indicating the safety system has detected a high level of refrigerant flow and stopped the injection, in applications one through six with the electrophysiology (ep) catheter with lot number 92896-21.The data files showed at least six applications were performed with this catheter on the date of the event.The data files also showed at least six more applications were performed with the electrophysiology (ep) catheter with lot number 92896-72 with system notice 50030 on all applications except application three which triggered system notice 50029, a similar notice indicating that the safety system detected a high level of refrigerant flow and stopped the injection.In conclusion, the decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician, and is the adverse event being reported.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|