Catalog Number 72203840 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that the spider was not working, the batteries were not charging on the hsc charger or the st clares hospital charger.The incident occurred before the procedure.There was no delay and a backup device was available to complete the procedure with no complications reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Updated, device received.The device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issues were observed.A functional evaluation revealed the device's indicator diode would not change to charging mode when a test battery was installed.The fuse was checked with an ohmmeter and was open.The complaint was confirmed and the root cause is an open circuit.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
|