Model Number PL579T
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/26/2020
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl579t - challenger ti-p ml-ligat.Clips 12 cartr.According to the complaint description, the cartridge fell in situ during surgery.Type of surgery: robot-assisted radical prostatectomy, all pieces have been recovered.An additional medical intervention was necessary.Additional information was not provided nor available.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).Involved components: pl536r - shaft compl.D:10mm l:370mm - lot: 62304844, pl520r - challenger ti-p handle - lot: unknown, pl536810 - jaw part protection for challeneger d 5 - lot: unknown.
 
Manufacturer Narrative
Investigation results: visual investigation: visually, the provided devices are in a used condition.The applier has been forwarded to the responsible q-coordinator of the production plant.It was found that the recommended maintenance date has been exceeded.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material, manufacturing- or design-related failure.
 
Event Description
The adverse event is filed under aag reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11066010
MDR Text Key223430602
Report Number9610612-2020-00961
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL579T
Device Catalogue NumberPL579T
Device Lot Number52625554
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/26/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received03/16/2021
Supplement Dates FDA Received04/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PL520R.; PL536810.; PL536R.; PL520R; PL536810; PL536R
Patient Outcome(s) Required Intervention;
-
-