It was reported that there was an issue with pl579t - challenger ti-p ml-ligat.Clips 12 cartr.According to the complaint description, the cartridge fell in situ during surgery.Type of surgery: robot-assisted radical prostatectomy, all pieces have been recovered.An additional medical intervention was necessary.Additional information was not provided nor available.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).Involved components: pl536r - shaft compl.D:10mm l:370mm - lot: 62304844, pl520r - challenger ti-p handle - lot: unknown, pl536810 - jaw part protection for challeneger d 5 - lot: unknown.
|
Investigation results: visual investigation: visually, the provided devices are in a used condition.The applier has been forwarded to the responsible q-coordinator of the production plant.It was found that the recommended maintenance date has been exceeded.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material, manufacturing- or design-related failure.
|