Model Number 24658
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2020
Event Type  malfunction  
Event Description
It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in a moderately tortuous and moderately calcified artery.A 4.00mm/1.5cm/140cm small peripheral cutting balloon was selected for use.During procedure, the target lesion did not dilate on the first dilation, so it was moved a little.Then, at second inflation, it was noted that the balloon ruptured immediately at 4atm.It seemed that there was a pinpoint calcification.Furthermore, there was visible pinhole noted on the device.The device was removed from the patient's body and the procedure was completed with a non boston scientific device.No patient complications were reported and patient's condition was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL PERIPHERAL CUTTING BALLOON
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11066031
MDR Text Key223443367
Report Number2134265-2020-18403
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/12/2023
Device Model Number24658
Device Catalogue Number24658
Device Lot Number0025506201
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial