H4: the devices were manufactured from 08/06/2020 ¿ 08/10/2020.H10: the initial report was for one device, however, two used samples were received for evaluation in a zip lock bag.Unaided eye visual inspection was done under normal room conditions which observed that the tube set tubing was received with detached spike found loose in the zip lock packaging of both samples.Functional testing could not be performed.The reported condition was verified on both samples.The cause of the condition could not be determined, however, the most likely cause was due to inadequate or lack of adhesive being applied in the manufacturing process at the spike connection to the tubing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|