MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
|
Back to Search Results |
|
Model Number BI70002000 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: other relevant device(s) are: product id: bi71000124, lot/serial: unknown.A medtronic representative went to the site to perform a system checkout.The issue was confirmed and the cable tray was found to be bent due to several components found within the gantry.The site stores the system in a supply closet with the gantry open.Due to the storage situation, components from the closet fall into the gantry and cause mechanical issues.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used during a spinal procedure.It was reported that when taking a spin, some loud noises were heard from the x-ray tube/detector and the spin was aborted.An alternate imagine system was used to complete the procedure.The procedure was delayed by less than one hour.There was no impact on patient outcome.After the case, the manufacturer representative opened the system and found an interbody box in the system.This was most likely impeding the tracks.
|
|