ARTHROCARE CORP. WEREWOLF FLOW 90 COBLATION WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 72290038
Device Problem Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during a rotator cuff repair, when the werewolf flow 90 coblation wand was being used a circuit failure came up in the controller.The wand then went back and forth between working normally and displaying the error message.The procedure was completed without significant delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: 1) there may have been an issue with one of the concomitant devices in use at the time of this complaint event.No containment or corrective actions are recommended at this time.There are no indications to suggest the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF FLOW 90 COBLATION WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11068999
MDR Text Key225009318
Report Number3006524618-2020-01143
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00885556647462
UDI-Public00885556647462
Combination Product (y/n)N
PMA/PMN Number
K183346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/19/2021
Device Model Number72290038
Device Catalogue Number72290038
Device Lot Number2041488
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received01/23/2021
Supplement Dates FDA Received01/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-