Model Number CD3365-40Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Information (3190)
Event Date 12/10/2020
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11070218
MDR Text Key223601822
Report Number2938836-2020-09696
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public05414734508292
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2016
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot Number4646193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0117-2017
Patient Sequence Number1
-
-