ARTHROCARE CORP. AMBIENT SUPER TURBOVAC 90 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number ASHA4250-01 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, when the super-tvac icw ifs ambient was connected to the console, it showed an alert on the screen reporting an error in the radio frequency, during an arthroscopy.The procedure was completed with another smith and nephew back up device (eliminator icw) and with a delay of less than or equal to 30 min.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10 h3, h6: the device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Please refer to the instructions for use for recommendations device usage.A review of risk management files found that the reported failure was documented appropriately.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: (1) previous use (2) disconnecting the wand from the controller after power has been turned on.(3) there may have been an issue with one of the concomitant devices in use at the time of this complaint event.There are no indications to suggest the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
|
Manufacturer Narrative
|
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.No visual issues were observed.A functional evaluation revealed the controller generated an e7 error when the wand was connected.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) previous use.(2) disconnecting the wand from the controller after power has been turned on.
|
|