• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIR LIQUIDE ISPAN SULFUR HEXAFLUORIDE (SF6) GAS; INTRAOCULAR GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIR LIQUIDE ISPAN SULFUR HEXAFLUORIDE (SF6) GAS; INTRAOCULAR GAS Back to Search Results
Catalog Number 8065797005
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2020
Event Type  malfunction  
Manufacturer Narrative
No sample or lot number information has been received by manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A surgeon reported that an ophthalmic gas was resorbing more quickly inside of the patient's eye than it used to.No patient harm was reported.Additional information has been requested.
 
Manufacturer Narrative
An ophthalmic gas sample was received and put through final inspection and testing.There was no problem found with the sample related to the reported event.The product met specifications upon sample testing.A review of the batch production record for the reported lot was performed and confirmed the product met specifications at the time of release.A review of complaints for the last 12 months did not indicate any additional related reports for this reported lot number of ophthalmic gas.No problem was found with the ophthalmic gas sample therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The previous report submitted for this event contained an error in h1: ¿summary report¿ was inadvertently selected.After a recent systems update, a system error caused the inadvertent additional selection of ¿summary report¿ in h.1.On a select number of reports.The error, which was limited only to the h.1.Field, was promptly identified and quickly rectified.Correction smdrs are being filed for the impacted reports.This smdr is correcting that error for this event.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional information received further clarified that 20% ophthalmic gas concentration mixed with room air was instilled into the patient's eye during their initial pars plana vitrectomy with endolaser surgery to treat a retinal detachment.The instilled bubble was only 55% of the anticipated size upon seven days post-op.The patient's current condition was reported as stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN SULFUR HEXAFLUORIDE (SF6) GAS
Type of Device
INTRAOCULAR GAS
Manufacturer (Section D)
AIR LIQUIDE
13140 ti blvd.
dallas TX 75243
MDR Report Key11071902
MDR Text Key224410334
Report Number1610287-2020-00040
Device Sequence Number1
Product Code LPO
Combination Product (y/n)N
PMA/PMN Number
P900067
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Catalogue Number8065797005
Device Lot Number925609
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2021
Initial Date Manufacturer Received 12/15/2020
Initial Date FDA Received12/24/2020
Supplement Dates Manufacturer Received12/28/2020
02/17/2021
03/03/2021
Supplement Dates FDA Received01/12/2021
03/05/2021
03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age64 YR
-
-