The procedure was not performed in accordance with the product instructions for use, as the stimulator was placed off-label at the craniofacial supra orbital nerve above the eyebrow.At a follow-up appointment, the implanting clinician prescribed antibiotics and scheduled an explant.The stimwave clinical representative (cr) did not attend the explant procedure as the patient was implanted out of state with a different cr.During the explant, the stimulator fractured.All fractured pieces of the stimulator were successfully removed without complication.There is no evidence that product did not meet specification, and the stimulator was used for treatment of pain.The reported complaint was confirmed.The stimulator was sterilized per specification, and sterile barriers were confirmed to be intact prior to implant.As the infection occurred more than one year after implant, product sterility is not the likely cause of the infection.The cause of the erosion and subsequent erosion/fracture is off-label use (user error).The craniofacial region has limited tissue to implant the stimulator at an adequate depth, which increases the likelihood of erosion if not placed and sutured appropriately.
|