• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TCH DEVICE&POWER SPLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TCH DEVICE&POWER SPLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801280
Device Problems Failure to Charge (1085); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, using a renasys tch, the pump was not charging.The device needed to be plugged in in order to work.It was reported to be slightly warmer than normal.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection reported no defects.The functional evaluation found the power supply was slightly warmer than normal establishing a relationship between the reported event and the device.Further evaluation reported the battery charged to 83%.The root cause is a failed power supply.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TCH DEVICE&POWER SPLY
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11073740
MDR Text Key223703375
Report Number8043484-2020-04392
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00040565126586
UDI-Public00040565126586
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66801280
Device Catalogue Number66801280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-