• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0636
Device Problem Positioning Failure (1158)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 12/02/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a capsule which failed to attach.The trocar needle was only able to deploy halfway.A retrieval net was used to remove the capsule.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.Lubrication was used to facilitate placement of the capsule and the delivery system and capsule will be returned for investigation.A repeat procedure was performed during the same event.
 
Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The capsule and delivery system were returned for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the capsule failed to attach to the patient's esophagus.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
MDR Report Key11074622
MDR Text Key223709429
Report Number9710107-2020-00746
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101369714
UDI-Public07290101369714
Combination Product (y/n)N
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/08/2021
Device Model NumberFGS-0636
Device Catalogue NumberFGS-0636
Device Lot Number50379F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2021
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received02/14/2021
Supplement Dates FDA Received03/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-