• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC. THERMOCOOL SMART TOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC. THERMOCOOL SMART TOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number D134804
Device Problems High Readings (2459); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Tamponade (2226)
Event Date 11/30/2020
Event Type  Death  
Manufacturer Narrative
The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device 30430125l number, and no internal action related to the complaint was found during the review.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.(b)(4).
 
Event Description
It was reported that a (b)(6)-year-old male patient with history of cardiomyopathy, underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and suffered cardiac tamponade (requiring pericardiocentesis and surgical intervention) and death.Physician mapped a right ventricle ectopic with the thermocool® smart touch® sf bi-directional navigation catheter.Mapping phase was successfully completed.After that, with the interest zone delimited in carto 3 system, physician decided to deliver radiofrequency to the right ventricle outflow tract (rvot) at 50w in power control mode with a smartablate generator.The physician was warned about delivering too high power but decided to continue.The force was monitored during the ablation phase and high contact force values (30g, 40g¿) were shown in the carto 3 screen.The physician was notified about the high force readings.The patient¿s blood pressure dropped, and cardiac tamponade was confirmed.Physician performed an emergency pericardiocentesis, but the patient did not recover and was transferred to surgery.An open-hear surgery was performed.Despite all efforts, the patient died.Physician causality opinion was not provided.Transseptal puncture was not done.No evidence of steam pop during the ablation was observed.The catheter flow was set at 17ml/min at 50 watts.The parameters for stability used with the visitag module were range: 3mm ¿ time: 3s, force over time (fot) 25% - 3grams.Tag size: 3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SMART TOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC.
33 technology drive
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
gabriel alfageme
31 technology dr.
irvine, CA 92618
9497898687
MDR Report Key11074818
MDR Text Key223702200
Report Number2029046-2020-01985
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010176
UDI-Public10846835010176
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/25/2021
Device Model NumberD134804
Device Catalogue NumberD134804
Device Lot Number30430125L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM.; UNK_SMARTABLATE GENERATOR.
Patient Outcome(s) Death; Required Intervention;
Patient Age67 YR
-
-