The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device 30430125l number, and no internal action related to the complaint was found during the review.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.(b)(4).
|
It was reported that a (b)(6)-year-old male patient with history of cardiomyopathy, underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and suffered cardiac tamponade (requiring pericardiocentesis and surgical intervention) and death.Physician mapped a right ventricle ectopic with the thermocool® smart touch® sf bi-directional navigation catheter.Mapping phase was successfully completed.After that, with the interest zone delimited in carto 3 system, physician decided to deliver radiofrequency to the right ventricle outflow tract (rvot) at 50w in power control mode with a smartablate generator.The physician was warned about delivering too high power but decided to continue.The force was monitored during the ablation phase and high contact force values (30g, 40g¿) were shown in the carto 3 screen.The physician was notified about the high force readings.The patient¿s blood pressure dropped, and cardiac tamponade was confirmed.Physician performed an emergency pericardiocentesis, but the patient did not recover and was transferred to surgery.An open-hear surgery was performed.Despite all efforts, the patient died.Physician causality opinion was not provided.Transseptal puncture was not done.No evidence of steam pop during the ablation was observed.The catheter flow was set at 17ml/min at 50 watts.The parameters for stability used with the visitag module were range: 3mm ¿ time: 3s, force over time (fot) 25% - 3grams.Tag size: 3.
|