Catalog Number FVM08100 |
Device Problems
Entrapment of Device (1212); Fracture (1260); Malposition of Device (2616)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The catalog number identified has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified.(expiry date: 03/2022).
|
|
Event Description
|
It was reported that during a stent placement procedure, the stent sheath was allegedly found to be ruptured upon deployment.It was further reported that the stent was stuck in the left iliac axis and it got dropped in the wrong place.There was no reported patient injury.
|
|
Event Description
|
It was reported that during a stent placement procedure, the stent sheath was allegedly found to be ruptured upon deployment.It was further reported that the stent was stuck in the left iliac axis and it got dropped in the wrong place.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a review of manufacturing records was not performed.No additional complaint has been previously reported for this lot number.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent graft delivery system was returned for evaluation.Based on the investigation of the returned delivery system, the reported fracture of the outer sheath of the system was confirmed.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling it was found that the instructions for use sufficiently address the potential risks.The instructions for use states "prior to stent graft deployment in tortuous anatomy, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy." regarding preparation of the device the instructions for use states that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment." h10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g5.H10: d4 (expiry date: 03/2022).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|
Search Alerts/Recalls
|
|